R O U B L E S H O O T I N G
Telemetry products use different text for their SpO
2
error messages
than non-telemetry products.
serves as a key for
interpreting the telemetry messages. Telemetry products also
display their error messages within ECGs display zone, so the
following ECG alarm messages take priority over other SpO
2
messages.
•
LEADS OFF
•
NOISY SIGNAL
•
ECG ALARMS SUSPENDED
ADAPTER DISCONNECTED — CHECK CONNECTION AT MODULE
•
The module does not detect an adapter cable connected to the
front panel. Check proper adapter cable connection.
•
If the message persists and the adapter cable is secure, replace
the adapter cable.
•
The channel will drop and the alarm will stop after approximately
10 seconds.
•
On remote view, there may be no audible alarm on the remote
monitor before the channel drops.
SENSOR DISCONNECTED — CHECK CONNECTION AT
ADAPTER CABLE
•
The module does not detect a sensor. Check the proper
connection of the sensor with adapter cable.
•
If the message persists, replace the sensor and/or the adapter
cable.
Table 5-1 Telemetry Message Equivalent
SpO
2
Message
Equivalent Telemetry SpO
2
Message
ADAPTER DISCONNECTED – CHECK CONNECTION AT MODULE
SPO2 UNAVAILABLE
FAULTY SENSOR – REPLACE SENSOR
SPO2 FAULTY SENSOR
SENSOR DISCONNECTED – CHECK CONNECTION AT ADAPTER
CABLE
SPO2 SENSOR DISCONNECTED
SENSOR OFF PATIENT – CHECK CONNECTION AT PATIENT
SPO2 SENSOR OFF PATIENT
INSUFFICIENT SIGNAL – REPOSITION OR REPLACE SENSOR
SPO2 INSUFFICIENT SIGNAL
AMBIENT LIGHT INTERFERENCE – COVER SENSOR AREA
SPO2 AMBIENT LIGHT INTF.
NOISY SIGNAL
SPO2 NOISY SIGNAL
LOW SIGNAL STRENGTH – REPOSITION OR REPLACE SENSOR
SPO2 INSUFFICIENT SIGNAL