5-10
VPH-G90E/G90U/G90M
5-6-2. Uploading Program (Emergency Mode)
1. When the 7-segment LED of the YB board changes as
follows when the AC is turned ON, the program upload
mode will forcibly be set due to the abnormal program
data inside. (Emergency mode)
Red : 88 (Approx. 0.5 seconds)
→
Orange : 88 (Approx.
1.0 seconds)
→
Green : 10 (blinks twice)
→
Green : 11
(communication standby)
2. Prepare to upload program.
Preparations : Set the S961 knob to the left.
(RS-232C
↔
RS-422A switch)
Connect the RS-232C controller.
3. Start the uploader software of the RS-232C controller,
and start uploading the program data according to the
proper procedure.
The 7-segment LED on the YB board changes as
follows;
Green : 11 (loding)
→
Green :12 (updating)
→
Green :
11
→
Green : 12
→
(repeated)
→
Green : ZF (Aoorix.
0.5 seconds)
→
Red : 88 (Approx. 0.5 seconds)
→
Orange : 88 (Approx. 1.0 seconds)
→
Off
4. Initialization ends after about 11 minutes.
→
If Green : 30 (Lit) lights up on the 7-segment
LED of the YB board without going off, it means
that the software is requesting initialization of the
user SRAM. Press S202 for about 5 seconds in
the Green : 30 (lit) state, and the 7-segment LED
should change as follows.
Green :30
→
31 (Approx. 0.1 seconds)
→
Off
End of intialization
*
Take note that initializing the user SRAM
erases all user data.
5. Turn on the power and check operations.
5-6-3. Creating Factory Data
To forcibly overwrite current factory data
1. Set S201-(1) to OFF in the AC OFF state.
→
Initial setting for entering the factory mode.
2. Perform AC ON, various adjustments, settings, data
saving (user area).
3.
In the AC ON state (or Standby state), set S201-(1) to
ON.
→
The data will automatically be overwritten when
the factory mode is exited.
4. The 7-segment LED changes from 38 lit to off.
*
Duration of 38 lit is changed in accordance with
data amount adjusted.
5. End of creating data
If data needs to be created again due to
program version upgrading and parts
replacement (IC334)
1. In the AC OFF state, set S201-(1) to OFF.
→
Initial setting for entering the factory mode.
4. The 7-segment LED changes as follows;
37 (approx. 90 seconds)
→
38 (approx. 40 seconds)
→
Off
5. End of creating data
2. Perform AC ON, various adjustments, settings, data
saving (user area).
3. In the AC ON state (or Standby state), set S201-(1) to
ON.
→
The data will automatically be overwritten when
the factory mode is exited.
1
ON
ALPS
2
3
4
1
ON
ALPS
2
3
4
1
ON
ALPS
2
3
4
1
ON
ALPS
2
3
4
Summary of Contents for VPH-G90E
Page 10: ......
Page 120: ......
Page 122: ...2 2 VPH G90E G90U G90M 2 1 3 Location 3 ME MB MC PB M L MA BA F BC NA BB ...
Page 190: ......
Page 204: ......
Page 210: ......
Page 224: ......
Page 337: ...9 1 9 1 VPH G90E G90U G90M SECTION 9 BLOCK DIAGRAMS ...
Page 363: ...2 3 4 5 A B C D E F G H 1 10 1 10 1 VPH G90J G90E G90U G90M SECTION 10 DIAGRAMS ...
Page 437: ...2 3 4 5 A B C D E F G H 1 10 75 10 75 VPH G90J G90E G90U G90M ...
Page 474: ...2 3 4 5 A B C D E F G H 1 10 112 10 112 VPH G90J G90E G90U G90M ...