4-3
PCS-1500/1500P
Symptom: The main power cannot be turned on at the
event of automatic receive call.
Symptom: The main power of monitor or TV is not turned
on when the main power is turned on.
4-1. Check Items Before Starting Self-diagnostics
YES
YES
NO
NO
NO
YES
NO
YES
NO
YES
SS-03
SS-02
0-0
0-0
Main power is turned on?
The main power
of the system is turned on
when the POWER ON
button of remote
commander is
pressed?
Automatic
receive call
of the line that is connected to
the port A can be performed
successfully
IF-744 board is used
Automatic
receive call of the line
can be performed successfully
when the IF-744 board
is replaced
Replace the IF-744 board
Turn on the main power
There is no problem
YES
YES
NO
NO
NO
YES
NO
YES
SS-04
0-0
The types of monitor
are supported by
PCS-1500/1500P?
Monitor setting
(System Configuration 3/3)
is correct?
Power cable of
the monitor is connected
and main power is
turned on?
Connection and
installation of IR repeater
are correct?
It cannot be controlled
Set it correctly
Turn on the main power
Connect and install it correctly