Never remove USB media when these indicators are
blinking green.
Supported USB devices
• Use Sony USB media.
• The connectors do not support all USB devices.
• USB hubs and devices with built-in hubs are not
supported.
Warning
Using this unit for medical purposes
The connectors on this unit are not isolated.
Always use USB-compatible external hard drives and
USB memory devices that operate solely on the
power supplied by the unit.
Connecting a device that operates on an AC power
supply may result in an influx of leakage currents
from the connected device, which may in turn result
in electric shocks to the patient and operator.
If use of such devices cannot be avoided, be sure to
connect an isolation transformer to the power supply
of the connected device or connect an isolator
between the connection cables to isolate the power
supply, and verify that the risk of electric shock from
the system has been sufficiently reduced to levels that
conform to IEC 60601-1 standards.
q
1
(on/standby) switch/indicator
Use this switch to turn on the unit and the front panel
display. Press this switch again to turn off the unit and
enter standby mode.
The indicator indicates the following.
The indicator will turn off about 15 seconds after the
main power switch on the rear panel is turned off
from the standby state.
Lit green
The recording area of the USB media is
full. Delete any unnecessary recorded
data, or insert different media.
Lit orange
An error has occurred.
Note
Tip
Indicator
Status
Off
The main power switch on the rear panel
is set to the
a
(off) position.
Lit orange
Standby mode. (The main power switch
on the rear panel is set to the
?
(on)
position.)
Lit green
The unit is turned on. (The
1
(on/
standby) switch was pressed to turn on
the unit.)