Theory of Operation
45
40-5648-51B
Sensors interface description
Motor speed detection
The digital sense signal is created by using a reflective
infrared optical sensor to sense a motor-coupled en
-
coder as the motor rotates. The microprocessor mea-
sures the period of the signal created by the encoder.
Syringe sensing description
The syringe sensing system senses the syringe bar-
rel size, senses the plunger end cap is secured by the
mechanical flippers, and the syringe flange (ears) are
located in the pump’s syringe flange locating device.
•
The circuit for sensing the plunger flippers is part
of the plunger printed circuit function.
Syringe size sensing
Syringe sensing is accomplished by using a potenti-
ometer which changes its resistance with the barrel
diameter. The potentiometer is excited with a preci-
sion 4.096v voltage reference and the potentiometer’s
output is monitored through channel 2 of the system
A/D.
•
The sensing system requires calibration to deter-
mine offset and gain. Calibration values are stored
in non-volatile (serial
EEPROM) memory.
Syringe flange (ear) sensing
Syringe flange (or ear) sensor is used to detect if the
syringe flange is installed in the flange-holder on the
side of the pump.
An infrared optical interrupter detects
when the
flange-locating device is in a valid position. A prop
-
erly loaded flange causes the optical path to become
uninterrupted allowing a direct path from the infrared
emitter to the detector.
•
The emitter is enabled by the signal
syringe_ear_
snsr_enbl
when set to logic low.
•
The sensor output,
syringe_ear_detect,
pro-
duces a high logic state when the flange is properly
loaded.
•
The state should be logic low anytime the syringe
ear (flange) is not installed or
syringe_ear_snsr_
enbl is inactive.
Summary of Contents for Medfusion 3000 Series
Page 1: ...s Medfusion 3000 Series Technical Service Manual ...
Page 2: ...Table of Contents ii 40 5648 51B ...
Page 20: ...Introduction 14 40 5648 51B ...
Page 32: ...Scheduled Maintenance 26 40 5648 51B ...
Page 80: ...Troubleshooting 74 40 5648 51B ...
Page 116: ...Calibration Adjustment 110 40 5648 51B ...
Page 160: ...Assembly Drawings Parts Lists 154 40 5648 51B Case bottom assembly Medfusion 3500BC ...
Page 162: ...Assembly Drawings Parts Lists 156 40 5648 51B Case top assembly w plunger ...
Page 164: ...Assembly Drawings Parts Lists 158 40 5648 51B Case top assembly without plunger ...
Page 172: ...Assembly Drawings Parts Lists 166 40 5648 51B Accessories parts list Poleclamp assembly ...
Page 182: ...Index 176 40 5648 51B ...
Page 183: ......