Calibration & Adjustment
105
40-5648-51B
Adjust contrast (voltage) setting
The factory default contrast voltage setting is 00BF
(however this value may vary for different LCD
display
manufactures). This is the nominal setting for most
lighting conditions and may be adjusted higher or lower.
1. From the
Calibration
menu, choose
Adjust
Contrast
.
1. CAL SIZE AND POSITION 5. CAL SYRINGE SIZE SENSOR
2. CAL FORCE SENSOR
6. ADJUST CONTRAST
3. CAL PRESSURE SENSOR
7. VIEW CALIBRATION DATA
4. CAL PLUNGER POSITION 8. SAVE CHANGES AND EXIT
PRESS THE NUMBER TO SELECT THE MODE
PRESS THE NUMBER TO SELECT
1
CAL SIZE AND POSITION
5
ADJUST CONTRAST
2
CAL FORCE SENSOR
6
VIEW CALIBRATION DATA
3
CAL PLUNGER POSITION
7
SAVE CHANGES AND EXIT
4
CAL SYRINGE SIZE SENSOR
2. Use the “
↑
” and “
↓
” keys to setup the contrast volt
-
age. “
↑
” increases contrast.
CONTRAST VOLTAGE:
00B6
ADJUST CONTRAST
-
PRESS ENTER TO CONTINUE
ENTER
↓
↑
CONTRAST VOLTAGE:
00B6
ADJUST CONTRAST
-
PRESS
<
ENTER
>
TO CONTINUE
ENTER
↓
↑
3. When ready, press
Enter
to accept the new contrast
voltage setting and return to the
Calibration
menu. You may calibrate another mode, choose
Save Changes and Exit
to exit calibration, or
press
â
to return to the
Biomed
menu.
Summary of Contents for Medfusion 3000 Series
Page 1: ...s Medfusion 3000 Series Technical Service Manual ...
Page 2: ...Table of Contents ii 40 5648 51B ...
Page 20: ...Introduction 14 40 5648 51B ...
Page 32: ...Scheduled Maintenance 26 40 5648 51B ...
Page 80: ...Troubleshooting 74 40 5648 51B ...
Page 116: ...Calibration Adjustment 110 40 5648 51B ...
Page 160: ...Assembly Drawings Parts Lists 154 40 5648 51B Case bottom assembly Medfusion 3500BC ...
Page 162: ...Assembly Drawings Parts Lists 156 40 5648 51B Case top assembly w plunger ...
Page 164: ...Assembly Drawings Parts Lists 158 40 5648 51B Case top assembly without plunger ...
Page 172: ...Assembly Drawings Parts Lists 166 40 5648 51B Accessories parts list Poleclamp assembly ...
Page 182: ...Index 176 40 5648 51B ...
Page 183: ......