Troubleshooting
61
40-5648-51B
System Failure Alarms
This section discusses the diagnosis of system failure
alarms in the Medfusion™ 3000 Series pump, and sug
-
gests possible causes and remedies for these alarms.
System failure alarm
Cause or remedy
a2d r
eferenCe
v
oltage
Bgnd t
eSt
Background self-test measured an invalid range on analog-to-digital con
-
verter reference voltage. Use
Biomed
>
Diagnostics
to check the volt
-
age. Replace the main board if invalid voltage found.
aCu P
oWer
S
troBe
f
ailure
Self-test has found that the ACU processor is not maintaining its power
strobe. If this failure repeats, then try full reset of main board – disconnect
AC power, open pump, disconnect battery, short super-cap through ~100
ohm load; then reassemble pump and retest. Replace main board if prob
-
lem persists.
aCu W
atChdog
f
ailure
Self-test has found that the ACU watchdog alarm signal active. This alarm
may occur if processor failed to pet watchdog within allotted time (e.g. a
software failure occurred). If problem recurs, replace the main board.
a
nalog
S
uPPly
Bgnd t
eSt
Background self-test measured an invalid range on analog supply volt
-
age. Use
Biomed
>
Diagnostics
to check the voltage. Replace the main
board if invalid voltage found.
a
ux
C
ontroller
u
nit
P
oSt
Power-up self-test has found that the ACU processor failed to shut down
motor current during the watchdog alarm test. If problem recurs, replace
the main board.
B
aCkground
CrC t
eSt
t
imeout
Software timing failure. A transient failure may have occurred in the soft
-
ware; if problem recurs, replace the main board.
B
aCkground
S
elf
t
eSt
t
imeout
Software timing failure. A transient failure may have occurred in the soft
-
ware; if problem recurs, replace the main board.
B
aCkuP
a
udiBle
a
larm
P
oSt
Power-up self-testing has found that the backup audible alarm (controlled
by ACU) is not working at correct power (current) or frequency. Replace
the alarm buzzer on the main board or replace the main board.
C
aliBration
r
equired
Calibration values stored in non-volatile serial EEPROM memory are
invalid. This message appears whenever a new main board is installed.
Calibrate all the sensors in the pump.
C
ontrol
k
ey
S
WitCh
Bgnd t
eSt
Key on keypad was found stuck on during continuous background self-
test. Use
Biomed
>
Diagnostics
to determine which key is stuck. If
stuck key is found, replace keypad.
C
ontrol
k
ey
S
WitCh
P
oSt
Key on keypad was found stuck on during power-on self-test. Use
Biomed
>
Diagnostics
to determine which key is stuck. If stuck key is found,
replace keypad.
C
ritiCal
d
ata
B
loCk
Bgnd t
eSt
Delivery settings are stored in a critical data block of RAM on the main
board. This message will occur if background self-test showed the criti
-
cal data block was corrupted (i.e. a RAM cyclic redundancy test failed).
Run power-on self test which executes extensive RAM testing – if problem
recurs, replace the main board.
Summary of Contents for Medfusion 3000 Series
Page 1: ...s Medfusion 3000 Series Technical Service Manual ...
Page 2: ...Table of Contents ii 40 5648 51B ...
Page 20: ...Introduction 14 40 5648 51B ...
Page 32: ...Scheduled Maintenance 26 40 5648 51B ...
Page 80: ...Troubleshooting 74 40 5648 51B ...
Page 116: ...Calibration Adjustment 110 40 5648 51B ...
Page 160: ...Assembly Drawings Parts Lists 154 40 5648 51B Case bottom assembly Medfusion 3500BC ...
Page 162: ...Assembly Drawings Parts Lists 156 40 5648 51B Case top assembly w plunger ...
Page 164: ...Assembly Drawings Parts Lists 158 40 5648 51B Case top assembly without plunger ...
Page 172: ...Assembly Drawings Parts Lists 166 40 5648 51B Accessories parts list Poleclamp assembly ...
Page 182: ...Index 176 40 5648 51B ...
Page 183: ......