Theory of Operation
48
40-5648-51B
Plunger flipper sensor function
Plunger flipper sensing requires each sensor detect
its respective flipper position. Each infrared optical
interrupter provides a high logic state when the flipper
is in a valid position determined by a loaded syringe
plunger. The transmitters for both photo-interrupters
are driven in series; however, the optical receivers are
read in parallel.
•
Output from each
photo-interrupter
is monitored
individually so each sensor is independently
checked.
•
When a syringe plunger is correctly loaded, both
plunger_snsr1_out
and
plunger_snsr2_out
are at a
logic high
.
•
The transistor Q1 allows the sensor output to be
tested by turning the transmitter from both sen
-
sors on and off. This is controlled by the signal
plunger_snsrs_enbl from the Main PCB.