Introduction
6
40-5648-51B
Button
When pump is paused
When pump is delivering
Alarm Silence (
í
)
Silences audible alarm. Allows opera-
tor to switch the display
backlight from
bright to dim or dim to bright.
Silences audible alarm. Allows operator to
switch the display backlight from bright to
dim or dim to bright.
Power (
ë
)
Push and hold to turn pump Off.
Push and release, then push and hold to turn
pump Off.
Menu buttons
Function is defined on the display.
Function is defined on the display.
Back (
â
)
Reverts to a previous step or level.
Reverts to a previous step or level if adjusting
settings.
Numbers & Decimal
Set number values or select options.
Set number values or select options.
Stop (
á
)
N/A
Stops delivery (pump remains On).
Start (
É
)
Starts delivery.
N/A
Bolus (
è
)
Begins priming after confirmation.
Begins delivery of programmed bolus after
confirmation.
Indicator
What it means
ò
- Alarm
The
Alarm
indicators (yellow or red) are on whenever the pump is in an alarm
condition. The specific details of each alarm are covered in the Troubleshooting
section later in this manual.
ô
- AC Line
The
AC Line
indicator (green) is On whenever the pump is connected to “mains”
line power. It is Off when the pump is not connected to an active AC line
.
õ
- Battery
The
Battery
indicator (green) blinks On & Off whenever the pump is operating on
internal battery power, and remains On when the battery is charging
.
ö
- Lock
The
Lock
indicator tells you the pump has been locked into its current operational
mode.
While this indicator is lighted, the keypad is locked and no changes can be made
to settings. Attempting to stop or change an infusion while locked will result in an
alarm and an advisory message
.
úù
-
Infusing
The
Infusing
indicators are 3 green lights, which illuminate in sequence right to
left when the pump is delivering fluid (the 4
th
LED indicator on the far left - on the
tubing - is not used at this time).
During intermittent delivery mode, a single Infusing indicator lights during the
time between infusions
.
When the pump is Off or stopped, the Infusing indicator is not lit.
Summary of Contents for Medfusion 3000 Series
Page 1: ...s Medfusion 3000 Series Technical Service Manual ...
Page 2: ...Table of Contents ii 40 5648 51B ...
Page 20: ...Introduction 14 40 5648 51B ...
Page 32: ...Scheduled Maintenance 26 40 5648 51B ...
Page 80: ...Troubleshooting 74 40 5648 51B ...
Page 116: ...Calibration Adjustment 110 40 5648 51B ...
Page 160: ...Assembly Drawings Parts Lists 154 40 5648 51B Case bottom assembly Medfusion 3500BC ...
Page 162: ...Assembly Drawings Parts Lists 156 40 5648 51B Case top assembly w plunger ...
Page 164: ...Assembly Drawings Parts Lists 158 40 5648 51B Case top assembly without plunger ...
Page 172: ...Assembly Drawings Parts Lists 166 40 5648 51B Accessories parts list Poleclamp assembly ...
Page 182: ...Index 176 40 5648 51B ...
Page 183: ......