Introduction
2
40-5648-51B
• While servicing the Medfusion
™
3000 Series infu
-
sion pump you should wear safety glasses as it
contains springs and other small parts which may
be a hazard.
Cautions
•
Avoid Organic Solvents:
Never use organic sol
-
vents (e.g., acetone), quarternary ammonia com
-
pounds, strong acids or strong bases to clean any
portion of the pump.
•
Not Waterproof:
The Medfusion™ 3000 Series
is “spray resistant” but not “water proof”. Never
spray cleaning or other fluids directly into open-
ings on the bottom of the pump. Never immerse
the pump in water or other fluids.
•
Avoid Spray Oils:
Never use light spray oils (e.g.,
WD40™) to clean or lubricate pump. These chemi
-
cals can damage the plastic of the pump.
•
Never Autoclave or Gas Sterilize:
Never sterilize
the pump in a steam autoclave or gas. Using an
autoclave or gas sterilization can seriously damage
the Medfusion™ 3000 Series pump and void the
warranty.
•
Disconnect AC Mains & External DC Power:
Always disconnect the pump from AC Mains and
from External DC power before disassembling the
pump for maintenance.
•
Handle Batteries with Care:
Always handle the
pump’s battery pack with care.
•
Don’t Over-tighten Screws:
Never over-tighten
any screws in the pump. Unless otherwise speci
-
fied, you should torque all screws to 60 in- oz
(0.42 Nm).
•
Battery Disposal:
Always dispose of exhausted
NiMH batteries in compliance with all pertinent
local, state, national, and international regulations.
If unsure of correct methods for compliance, you
may return battery packs to Smiths Medical for
recycling.
•
Keypad is NOT Flexible:
Whenever handling the
keypad, always ensure it remains flat. Bending the
keypad can damage keys or break LED contacts.
Summary of Contents for Medfusion 3000 Series
Page 1: ...s Medfusion 3000 Series Technical Service Manual ...
Page 2: ...Table of Contents ii 40 5648 51B ...
Page 20: ...Introduction 14 40 5648 51B ...
Page 32: ...Scheduled Maintenance 26 40 5648 51B ...
Page 80: ...Troubleshooting 74 40 5648 51B ...
Page 116: ...Calibration Adjustment 110 40 5648 51B ...
Page 160: ...Assembly Drawings Parts Lists 154 40 5648 51B Case bottom assembly Medfusion 3500BC ...
Page 162: ...Assembly Drawings Parts Lists 156 40 5648 51B Case top assembly w plunger ...
Page 164: ...Assembly Drawings Parts Lists 158 40 5648 51B Case top assembly without plunger ...
Page 172: ...Assembly Drawings Parts Lists 166 40 5648 51B Accessories parts list Poleclamp assembly ...
Page 182: ...Index 176 40 5648 51B ...
Page 183: ......