Calibration & Adjustment
96
40-5648-51B
Below is table which summarizes the options available
for calibration.
Calibration Option
What it Controls
Calibrate (Syringe) Size and Position
Performs concurrent adjustment of syringe size & position sensors
settings. This requires Smiths Medical “calibration slugs”.
Calibrate Force Sensor
Performs adjustment of force sensor settings.
Calibrate Pressure Sensor
This is a future option of the Medfusion™ 3000 Series pumps.
Calibrate Plunger Position Sensor
Performs adjustment of plunger position sensor settings.
Calibrate Syringe Size Sensor
Performs adjustment of syringe size sensor setting.
Adjust Contrast
Varies the onscreen contrast of the LCD. The pump is shipped with
a “factory average” default. You may set it above or below for “com
-
fortable” viewing.
View Calibration Data
Displays the “current” calibration data settings.
Save Changes and Exit
Saves the new calibration settings and then returns to the main
Biomed
menu.
Summary of Contents for Medfusion 3000 Series
Page 1: ...s Medfusion 3000 Series Technical Service Manual ...
Page 2: ...Table of Contents ii 40 5648 51B ...
Page 20: ...Introduction 14 40 5648 51B ...
Page 32: ...Scheduled Maintenance 26 40 5648 51B ...
Page 80: ...Troubleshooting 74 40 5648 51B ...
Page 116: ...Calibration Adjustment 110 40 5648 51B ...
Page 160: ...Assembly Drawings Parts Lists 154 40 5648 51B Case bottom assembly Medfusion 3500BC ...
Page 162: ...Assembly Drawings Parts Lists 156 40 5648 51B Case top assembly w plunger ...
Page 164: ...Assembly Drawings Parts Lists 158 40 5648 51B Case top assembly without plunger ...
Page 172: ...Assembly Drawings Parts Lists 166 40 5648 51B Accessories parts list Poleclamp assembly ...
Page 182: ...Index 176 40 5648 51B ...
Page 183: ......