Chapter 5 Product Operation
5.3 Operation Screen
HRZD Series
5-36
5.3.23 CH1(2) analog input setting screen
Figure 5-53 “CH1(2) Analog Input Setting Screen 1”
Table 5-26 “CH1(2) Analog Input Setting Screen 1”
No.
Item
Setting range
Descriptions
Factory
default
1 [TYPE]
key
DI SP, N/A, TEMP
SP, FLOW SP
Used to specify the type of analog input signal.
DI SP is displayed only if the DI control kit (optional) is
provided.
TEMP SP
2 SCALING
―――
Used to specify the scaling of analog input signal.
-
3
[V Lo] key
―――
Used to display the “Ten-key screen”.
0 V
4
[V Hi] key
―――
Used to display the “Ten-key screen”.
10 V
5 [Enter]
key
―――
Used to display the “MESSAGE screen” to accept the
setting.
-
6 [Initial]
key
―――
Used to display the “MESSAGE screen” to initialize
the settings of TYPE and SCALING.
-
7 [Cancel]
key
―――
Used to cancel the setting.
-
8 [Return]
key
―――
Used to display the “CH1(2)DIO communication
setting screen 1” on page 5-32.
-
[Tips]
The setting range of items is displayed on the “Ten-key screen” with the
touch of the [V Lo] or [V Hi] key.
A scaling must be specified to cause V Hi to be larger than V Lo (V Lo<V Hi).