(on)
(on)
(on)
(on)
(on)
(on)
(on)
(on)
(on)
(on)
Temp:Chambr_elec
xx.x°C
Conductivity
xxx µS/cm
Reserve
xx mbar
Temp_Preh_Cham.
xx.x°C
Pressure_Chamber
xxxx mbar
Temp_Chambr_Dis
2
Temp_Chambr_Dis
1
I1: Door contact
closed
I7: W. water
Swimmer closed
I6: Aqua dem
swimmer closed
I5: Door locking dev.
open
I4: Aqua dem, current
open
I3: Reserve open
I2: OH protector HT
closed
O1: Steam-generator
x.x bar xx°C (off)
O7: SV Press. rel.1
x.x bar xx°C (off)
O6: SV Vacuum 1/2
x.x bar xx°C (off)
O5: Pump/SV aqua dem
x.x bar xx°C (off)
O4: Fan
x.x bar xx°C(off)
O3: Vacuum-pump
x.x bar xx°C (off)
O2: Pre-heating
x.x bar xx°C (off)
O10: Door lock dev.
x.x bar xx°C (off)
O9: SV Press. rel. 2
x.x bar xx°C (off)
O8: SV Ventilation
x.x bar xx°C (off)
Main- and Submenues can be changed:
forward and with backwards scrolling
and always leaving the loop
Diagnosis
Analog-Inputs
Diagnosis
Digital-Inputs
Diagnosis
Digital-Outputs
Diagnosis
Keyboard
Diagnosis
Self-Test
Diagnosis
Malfunction-mem
Diagnosis
Maintenance-data
Diagnosis
Statistics
Diagnosis
Delete Backup
Diagnosis
Backup Parameter
Diagnosis
Memory init.
Diagnosis
Version
hh:mm:ss
x.xx bar xx°C
Page 1
Program-cycle / Diagnosis-program
Nitram Dac Professional
Service: dd.mm.yy
run cycles: xx
Delete Backup
Sure?
Backup Parameter
Sure?
Memory init.
Sure?
Malfunction-mem
print?
Service: dd.mm.yy
run cycles: 0
Maintenance-data
Update?
Total: xxxx
succesfull.: xxxx
xxxxklav XX-X typ-
progvers spr parvers
Keyboard
Nr.: 0
Keyboard
Nr.: 10
Keyboard
Nr.: 20
Keyboard
Nr.: 40
Keyboard
Nr.: 80
Self-Test
Signal
Sirona-GmbH
Gerätename xxx
Self-Test
Memory-Test OK
Self-Test
Battery OK
Printer is
not ready
Press both buttons
and hold ...
... after power is
switched on.
S
P
+
-
-
+
Buttons:"Continue" in the Menu ing diagnosis-rogram
Button choose program:"Enter/Quit/Input" in the diagnosso-program
Button Start/Stop: "Escape/Leave/Interrupt" in the diagnosis-program
-
+
-
+
-
+
-
+
P
P
P
P
P
P
P
P
P
P
P
P
P
-
+
S
S
P
P
P
P
S
-
+
P
P
P
P
P
P
P
P
S
S
S
S
Vacuklav 30-B
V30-3.16 D01.13
Symbol
Meaning
Example
typ
Type
V30
xxxxklav XX-X
Name
Vacukla 30-B
progvers
Programversion
3.16
parvers
Parametervers.
01.13
spr
Language
D
-
+ P
-
+ P
-
+ P
P
Neutral position
-
+
P
S
S
S
S
09/21/2006
59/61
Servicemanual DAC Professional
Summary of Contents for DAC PROFESSIONAL
Page 1: ...b a PROFESSIONAL p j...
Page 7: ...Figure 4 Back view 09 21 2006 5 61 Servicemanual DAC Professional...
Page 56: ...Sirona DAC Professional Process 09 21 2006 54 61 Servicemanual DAC Professional...
Page 60: ...Valve diagram 09 21 2006 58 61 Servicemanual DAC Professional...
Page 64: ......