Bowie & Dick test
1. Place the complete package in the autoclave.
The unused indicator shows the colour red.
2. open the package.
3. If the indicator has only been changed partly the
penetration is not satisfactory.
4. Only if the colour of the indicator is uniform the
penetration is satisfactory.
According to EN13060 this is a PCD Process Challenge Device.
Bowie&Dick test set:
mat no.: 58 92 034
Helix Test Set:
mat. no.: 58 92 042
09/21/2006
57/61
Servicemanual DAC Professional
Summary of Contents for DAC PROFESSIONAL
Page 1: ...b a PROFESSIONAL p j...
Page 7: ...Figure 4 Back view 09 21 2006 5 61 Servicemanual DAC Professional...
Page 56: ...Sirona DAC Professional Process 09 21 2006 54 61 Servicemanual DAC Professional...
Page 60: ...Valve diagram 09 21 2006 58 61 Servicemanual DAC Professional...
Page 64: ......