Fault diagnosis for
Fault 14:
⇒
The amount of water supplied by the feed-water dosing unit
33-00
is not enough; test as follows:
•
Place the end of the feed-water hose
41-00-A
(behind the flow monitor) into a measuring beaker.
•
Switch on “A5: Pump/SV feed pump” (
AC OUT 5
) in the digital outputs of the
Diagnosis Program
.
•
Check the feed rate for the water: it must be 1 … 1.5 l/min.
•
If necessary, replace the defective feed-water pump including the solenoid valve
33-00
.
⇒
Test the signal processing at
DIN 4
of the main circuit board
08-00
, and check the cable connection to the flow sensor
41-00
as follows:
•
Check display “E4: Aqua dem, current” in the digital outputs of the
Diagnosis Program
.
•
First disconnect both of the plugs of the reed contact
41-00-B
, and then make a short-circuit by placing a jumper at the digital output. Then, if
necessary, pull out the plug of the digital inputs on the main circuit board
08-00
, and directly jumper
DIN 4
. Observe whether the display is
“open” or “closed”.
•
If it is defective, replace the main circuit board
08-00
. Repair the cable connection is required.
⇒
The flow sensor
41-00
at
DIN 4
is either defective, or not adjusted properly:
•
If the switching point is not correctly set, see the
Adjustment Instructions
.
•
The switching piston with magnet (
41-02
) may be jammed.
•
The reed contact may be defective (
41-01
)
•
Adjust the flow sensor
41-00
if necessary. If required, repair it or replace it.
09/21/2006
31/61
Servicemanual DAC Professional
Summary of Contents for DAC PROFESSIONAL
Page 1: ...b a PROFESSIONAL p j...
Page 7: ...Figure 4 Back view 09 21 2006 5 61 Servicemanual DAC Professional...
Page 56: ...Sirona DAC Professional Process 09 21 2006 54 61 Servicemanual DAC Professional...
Page 60: ...Valve diagram 09 21 2006 58 61 Servicemanual DAC Professional...
Page 64: ......