Fault-diagnosis DAC Professional
Fault 36: Sterilization PU
FAULT 36: Sterilization PU
Triggered by:
The calculated temperature required to produce saturated steam on the basis of the measured pressure is below
the minimum temperatures T1 to T5
Causes / Remedy
Operating fault/Checks by personnel:
•
maximum load exceeded
Appliance fault / Repair by technical service personnel
►
Pressure drop during the sterilisation phase
•
Leaks
Ventilation or Pressure release solenoid valves, spring-operated safety valve
•
"Overheating protection - steam generator" trips
The capillary control unit "Overheating protection - steam generator" interrupts the heating power during
the sterilization phase
•
Capillary tube level controller (digital input DIN 2) faulty
Controller does not open in the event of overheating (Water shortage), water shortage is therefore
not recognised and the capillary tube controller for overheating protection switches off.
(see also "Fault 2 Steam generator")
Observe level control in condensation return pipe, Exchange steam generator
•
Capillary Overheating protection control for Steam generator faulty
Controller switches too early - Exchange steam generator
Observe level control in condensation return pipe, Exchange steam generator.
►
Pressure sensor signal fluctuates
Check on the basis of the print-out records:
Print-out shows fluctuating pressure values, dipping below P 6 – P10
In the print-out, the values from the temperature sensor "Temp_chambr_elec " (AIN1) and from
temperature sensor "Temp_Chambr_Dis1/2" (AIN2) correspond.
pressure sensor: white = Signal, green= grd, brown = 24 VDC
signal voltage at 1 bar = 2.5 V DC, max. 10 V DC at 4bar
•
Exchange pressure signal
•
If malfunction report continues: Exchange printed circuit board
►
"Poor control behaviour of steam generator"
Poor heat transfer from the heating coil of the steam generator, in particular during the fluctuation on transition
from heating up to the sterilization phase it is possible that the temperature falls below the theoretical
temperature required for sterilization for short periods.
Increase settings for P6-P10, Z24 and Z25,
Exchange steam generator
09/21/2006
48/61
Servicemanual DAC Professional
Summary of Contents for DAC PROFESSIONAL
Page 1: ...b a PROFESSIONAL p j...
Page 7: ...Figure 4 Back view 09 21 2006 5 61 Servicemanual DAC Professional...
Page 56: ...Sirona DAC Professional Process 09 21 2006 54 61 Servicemanual DAC Professional...
Page 60: ...Valve diagram 09 21 2006 58 61 Servicemanual DAC Professional...
Page 64: ......