Required Space:
Before setting up the autoclave you should make sure that the following requirements are met:
The DAC Professional is not designed to be installed in a cupboard or enclosed space. In order to
ensure adequate ventilation, there should be at least 10 cm clearance to the sides (for dimensions
see Figure1), and 10 cm clearance at the rear for connections.
Above the autoclave there must be sufficient space to allow access to the built-in water storage tank
(1) and for a good circulation of air.
The DAC Professional operates with a blower (8) on the rear of the unit, as part of the cooling
system.
The functioning and the service life of the unit can be impaired if the blower (8) is not allowed to
freely remove the heat from the unit.
Important:
Before installing the unit, make sure that there is sufficient free circulation of air
around the unit.
A plane, horizontal surface is required which is sufficiently stable to support the autoclave
(unloaded weight 43 kg).
Make sure that the inside of the chamber is in level and then turn the front feet 3 times down (see
figure 2).
Figure 2
The inside of the chamber
should be in level and then
the front feet are turned 3
turns down.
09/21/2006
3/61
Servicemanual DAC Professional