82 (104)
Revision 4
Table 31: Example values for parameter 11 to 14
No. Parameter
Calibration variable
Parameter value
1.
(11) Integer part of x-intercept
𝜶𝜶
𝜶𝜶 = −𝟖𝟖. 625
−
08
2.
(12) Decimal part of x-intercept
𝜶𝜶
𝜶𝜶 = −8. 𝟔𝟔𝟔𝟔𝟔𝟔
625
3.
(13) Integer part of slope
𝜷𝜷
𝜷𝜷 = 𝟏𝟏. 625
001
4.
(14) Decimal part of slope
𝜷𝜷
𝜷𝜷 = 1. 𝟔𝟔𝟔𝟔𝟔𝟔
625
Parameter 11: Integer part of x-intercept
𝜶𝜶
Parameter 11 is used to set the calibration value for the integer part of x-intercept
𝜶𝜶
when the device is
calibrated to display measurement results in mg/m
3
.
Table 32: Parameter 11 values
Value range (mg/m
3
) Description
Default
−
99 ... 099
Value for the integer part of x-intercept
𝜶𝜶
.
000
Parameter 12: Decimal part of x-intercept
𝜶𝜶
Parameter 12 is used to set the calibration value for the decimal part of x-intercept
𝜶𝜶
when the device is
calibrated to display measurement results in mg/m
3
.
Table 33: Parameter 12 values
Value range (mg/m
3
) Description
Default
000 ... 099
Value for the decimal part of x-intercept
𝜶𝜶
.
000
Parameter 13: Integer part of slope
𝜷𝜷
Parameter 13 is used to set the calibration value for the integer part of slope
𝜷𝜷
when the device is
calibrated to display measurement results in mg/m
3
.
Table 34: Parameter 13 values
Value range
(mg/m
3
/mA)
Description
Default
000 ... 099
Value for the integer part of slope
𝜷𝜷
.
000
Parameter 14: Decimal part of slope
𝜷𝜷
Parameter 14 is used to set the calibration value for the decimal part of slope
𝜷𝜷
when the device is
calibrated to display measurement results in mg/m
3
.