See Table 30 for descriptions of the parameter values:
Table 30: Parameter 10 values
Value
Description
001
Reset all parameters to default factory values.
002
Enable mg/m
3
calibration.
003
Disable mg/m
3
calibration.
Reset to default factory values
To reset the device to the default factory settings:
1.
Press K0 (
+
) key to enter MAINTENANCE mode.
2.
Navigate to parameter 10.
3.
Enter parameter value 001 and press K2 (
↩
) key to save the value.
4.
Press K3 (
←)
key to exit and return to normal operation.
All device parameters are now overwritten with the default factory values.
Resetting the device to factory settings with command parameter 10 rewrites all
parameter values, including the ones set with Auto Setup.
Enable mg/m
3
calibration
To enable displaying the measurement results in mg/m
3
:
1.
Press K0 (
+
) key to enter MAINTENANCE mode.
2.
Enter the values
𝜶𝜶
and
𝜷𝜷
of the calibration equation
𝒚𝒚
𝒊𝒊
= 𝜶𝜶 + 𝜷𝜷𝒙𝒙
𝒊𝒊
to parameters 11, 12, 13 and 14.
After setting the value for each parameter, press K2 (
↩
) key to save the new value.
See section
Parameters 11 to 14: Displaying measurement results in mg/m
the calibration equation and individual calibration parameters.
3.
Navigate to parameter 10.
4.
Enter parameter value 002 and press K2 (
↩
) key to save the value.
5.
Navigate to parameter 1.
6.
Enter parameter value 002 and press K2 (
↩
) key to save the value.
7.
Press K3 (
←)
key to exit and return to normal operation.
The measurement results are now displayed in mg/m
3
.