Operating Manual - English
Triax QUATTRO
Copyright Silca 2011
- 51 -
5.10.1
CLAMP CALIBRATION
F1
F2
F3
F1
to access the list of clamps provided by Silca.
Select the clamp to be calibrated.
Press ENTER.
F4
Press
START
to proceed with automatic calibration.
Press
ENTER,
the display will show the message on the right here.
Make sure tracer point U103 is fitted.
N.B.
Carefully clean the clamp and calibration plate
ATTENTION!
ELIMINATE SWARF FROM
CALIBRATION PLATE AND
CLAMP
Press START to proceed with calibration.
When calibration is finished the display shows the differences in the
values read.
Calibration will be accepted only if the size of the variations is between
+40 and -40 hundredths of millimetre.
Press ENTER to confirm.