Switching On the Calibrator
Switch on the power switch.
"Loading ...“ will be displayed. Then the start data will be displayed in the info window.
TP 38 165 / TP 38 165 E (Setting range ... -35 °C ... 165 °C)
TP 38 650 / TP 38 650 E (Setting range ... 40 °C ... 650 °C)
Below this, the firmware version, the date of the last calibration and the internally stored serial number are
displayed.
Displayed for 5 s only
Fig. 3: Info windows with start datas
6.3 Description of the control mode
After approx. 5 s, the calibrator will automatically switch to
Control mode
and display the current
block
temperature
. The
temperature set-point
is displayed below it. The two displays are permanently visible.
Fig. 4: Display in control mode
In addition, five other displays may appear depending on the equipment and selection. The figure above
shows the maximum scope of displays.
ext. Ref
displays the value of the external reference.
block temp 399.80°C
set temp 400.00°C
ext.ref. 350.80°C
PT100-3 350.55°C
probe PT100-3
Min. value 50.00°C
Max. value 500.78°C
control OFF 13:24:56
control