4 Structure of the temperature calibrator
SIKA Dr.Siebert & Kühn GmbH & Co.KG
4.8 Accessories (included in the scope of delivery)
Power cable
A power cable with a 90 ° angled inlet connector for non-heating apparatuses for the power supply is included
with every calibrator«TITLE». Depending on the country the calibrator is delivered to, the power cable is fitted
with the locally used power plug.
Important!
Operate the TP 38… / TP 38… E exclusively with the power cable provided. The power cable connector is on
the bottom of the calibrator, i.e. when using a different type of cable, the calibrator may topple over.
Insert removal tool
An insert removal tool for simple and easy removal and insertion of the adapter insert into the calibrator block
is included in the standard set of accessories of every temperature calibrator«TITLE» of the TP 38… / TP 38…
E series.
Test certificate
Each calibrator has been tested ex works and is delivered with a test certificate.
4.9 Optional Accessories
Adaptor inserts
Each calibrator features a block bore with a diameter of 28 mm. Adaptor inserts must be used to guarantee
that the heat transfer between the thermometer to be checked and the metal block is as small as possible. If
the diameters of your thermometers differ, please order the desired adaptor insert with an internal bore
matching the diameter of your thermometer plus 0.5 mm. Inserts have outside dimensions of 28 mm in Ø x a
length of 150 mm and are made of brass (scale-free special alloy).
Special adaptor inserts
Special adaptor inserts can be customized in various designs. Inserts with multiple bores are also possible.
Limits are placed only by the mechanical machinability of the insert blank.
If you require multiple inserts, please take the following into consideration:
if possible, the bores should be arranged on a pitch circle to ensure uniform temperature distribution
for product engineering reasons, there must be a minimum distance between the bores and to the edge of
the insert. As a rule, this distance should not be less than 2 mm.
the bore diameter should be larger than the thermometer to be tested by at least 0.5 mm.
Transport case
We offer a robust transport case for your calibrator as an option. We recommend the use of a transport case, if
you frequently use the calibrator for on-site calibrations and thus need to transport it often. The case does not
only protect your calibrator against environmental influences such as dust, dirt and humidity but also against
mechanical damage that may occur due to impacts and vibrations.
Data cable
A data cable for connecting the calibrator to the RS 232 interface can be ordered from SIKA.