SIKA Dr.Siebert & Kühn GmbH & Co.KG
Display filter
Below the "display filter“ line you can set the display attenuation. The display filter is calculated through
averaging. The values are entered using the block of 14 keys. The values entered are confirmed using
the ENTER key. The Clear key can be used to delete the selected numeric value. A value of "1“ means
that display attenuation is not performed. You can select a value between 1 and 99. 10 is preset ex
works. The set display filter is displayed in the Control, Calibration, Switch test, and Level test modes.
6.5.8 System value menu
After selecting the "systemvalue “ menu item in Select mode using the cursor key and confirming your
selection with the Select key, the "Select System" submenu will be displayed.
The settings for the calibration system are made in this menu. Make your selection using the cursor keys.
Activate your selection using the Select key.
Fig. 17: Submenu select system
Setting options
The temperature values and percentages are entered using the block of 14 keys. The values entered are
confirmed using the ENTER key. The Clear key can be used to delete the selected numeric value.
Language
In the first line you can select the desired language.
The selected language is displayed in the Control, Calibration, Switch test, Step test and Select modes.
Time/Date
This is where you can refresh the time and date.
The set time is displayed in the Control, Calibration, Switch test, step test and Select modes on the right
hand side of the status bar.
Brightness and contrast
Change of the display brightness and contrast in percent. The settings take effect in the Control,
Calibration, Switch test, Step test and Select modes. The contrast can be adjusted only by the factory.
Alarm
Here you can enter a temperature, at which the calibrator displays an alarm but continues performing its
control functions. The "TEMP-ALARM" alarm message is displayed in the Control, Calibration, Switch
test, Step test and Select modes on the left hand side of the status bar.
Switch-off
Here you can enter a temperature, at which the calibrator displays an alarm and switches off the control
unit. After switch-off, the program will branch to Control mode. The "TEMP-ALARM" alarm message is
displayed in the Control mode on the left hand side of the status bar. The calibrator will remain in
Control mode and can be reactivated via the Control key.