SIKA Dr.Siebert & Kühn GmbH & Co.KG
6.5.6 Step function menu
Important!
Take the settings for the step test only in „
CONTROL OFF"
mode.
After selecting the "step function" menu item in Select mode using the cursor key and confirming your
selection with the Select key, the "select steps" submenu will be displayed.
The settings for a step test are made in this menu. Make your selection using the cursor keys.
Activate/deactivate your selection using the Select key.
If the check or calibration is enabled with "upward" or "down" , operation will be initiated in step test mode.
Fig. 14: Submenu „select steps“
Switching ON/OFF
In the first line you can switch the step test ON respectively OFF.
Switching ON: Activation of field “upward” and/or of field “down”. (
√
characterizes the possibility of
multiple choice)
Switching OFF: field “OFF” („
●
“ characterizes the individual choice)
Off
Switches
the
level
test
off
upward
Levels are passed in ascending order
down
Levels are passed in descending order
upward/down
Levels are passed in ascending and descending order
Setting Options
Steps
You can enable/disable levels 1...6. In this case, "v “ identifies multiple selection options. The temperature
value is entered using the block of 14 keys. The temperature values entered are confirmed using the
ENTER key. The Clear key can be used to delete the selected numeric value.
Tolerance
Stability limits of the programmed temperature level
Gradient
defined heating and cooling of the calibrator
Time
Length of the dwell time on the programmed temperature level
Cycles
Number of repetitions