SIKA Dr.Siebert & Kühn GmbH & Co.KG
6.5.2 Select TC menu
This menu item is displayed if an internal measuring instrument is available and switched on.
After selecting the "calibr TC" menu item in Select mode using the cursor key and confirming your selection
with the Select key, the "Select TC" submenu will be displayed.
The settings for a connected thermocouple are made in this menu. Make your selection using the cursor keys.
Activate your selection using the Select key.
If the check or calibration of a TC thermocouple is enabled with "ON", operation will be initiated in "calibr TC"
mode.
Fig. 11: Submenu „select TC“
Switching ON/OFF
In the second line you can switch the thermocouple check on or off. In the third line you can select the type of
probe.
Probe Types
TCK
Thermocouple type K (NiCr-NiAl)
TCR
Thermocouple type R (Pt13Rh – Pt)
TCL
Thermocouple type L
(Fe-CuNi)
TCE
Thermocouple type E (NiCr-CuNi)
TCS
Thermocouple type S (Pt10Rh-Pt)
TCN
Thermocouple type N (NiCrSi – NiSiMg)
TCB
Thermocouple type B (Pt30Rh-Pt6Rh)
TCJ
Thermocouple type J
(Fe-CuNi)
TCT
Thermocouple type T (Cu-CuNi)
TCU
Thermocouple type U (Cu-CuNi)
Connection
Below the "Connection” line you have to select, whether the probe is connected to a 4 mm terminals or a TC-
plug.