SIKA Dr.Siebert & Kühn GmbH & Co.KG
Select RTD
Select TC
Select mA
Select Schalt
Select dislay
Select Logger
Select steps
Select system
Select info
Step
test
Select
steps
Display
value
Select
dislay
System value
Select system
Info
Select
Info
Submenu for resistance thermometers (RTD)
Submenu for thermocouples (TC)
Submenu for mA transmitters (mA)
Submenu for switch test (switch)
Submenu for display values in the display area (display)
Submenu for DataLogger
Submenu for steps (steps)
Submenu for system settings (system)
Submenu for information windows (info)
Display Area
In this area, you can read all the important data such as the block and temperature set-point as well as define
settings required for the calibration in the submenus on the respective screen pages.
Status Bar
The status bar displays the statuses of the calibrator and the time. The calibrator may have one of five
different statuses:
Heating
when the temperature in the block is increased
Cooling
when the block is cooled
Stable
when the temperature in the block has set to a constant value
(for stability, refer to the technical data)
Temp-alarm
when an alarm value set in the
select system
submenu has been exceeded
The alarm is signaled only, the control unit will continue operating.
when a switch-off value set in the
select system
submenu has been exceeded
The alarm is signaled. The control unit is switched off and the program
automatically branches back to the Control mode. Once the value has dropped
below the switch-off value, the control unit can be reactivated via the Control key.
Control Off
when the controller of the calibrator was manually switched off via the Control
key. The controller cannot be switched off manually while level and switch tests
are
being
performed.
6.1.2 Description of the Cursor Keys / Select Key
The cursor keys are used to select options on the display, e.g. a probe in Select
mode. The selected value will be displayed in blue letters on a white field.
The Select key is used to confirm your selection.
Using the Select key, you can call the Select mode including the associated
submenus even from the standard calibration mode. (please also refer to chapter
6.1.4)