SIKA Dr.Siebert & Kühn GmbH & Co.KG
6.5.4 Select switch menu
This menu item is displayed if an internal measuring instrument is available and switched on.
After selecting the "switch test menu item in Select mode using the cursor key and confirming your selection
with the Select key, the "select switch” submenu will be displayed.
The settings for a temperature switch test are made in this menu. Make your selection using the cursor keys.
Activate your selection using the Select key.
If the check or calibration of a temperature switch is enabled with "ON", operation will be initiated in switch test
mode.
Fig. 13: Submenu „select switch“
Switching test ON/OFF
In the second line you can switch the switch test on or off. In the third line you can set the conditions for the
temperature switch program.
Setting options
The temperature value is entered using the block of 14 keys. The temperature values entered are confirmed
using the ENTER key. The Clear key can be used to delete the selected numeric value.
change
Here you define the start and end point of the test. At this temperature, the
switch can be placed into the calibrator and removed from it.
Tmin/Tmax
This is where you define the temperature range, in which the switch test is
performed with a preset gradient.
gradient
Here you can define the gradient (
K
elvin
p
er
m
inute), with which the calibrator
passes through the selected temperature range.