Instruction Manual ScrubberGuard
Servicing
74
14830E/1
9.2.3
Cleaning the water tank
The following describes cleaning the water tank:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Remove the casing on the system so that the
water tank (arrow) is easily accessible.
2.
Close the sample feed and drain the water
tank by pressing the
Drain tank
button (Sec-
tion 7.1).
The pump may only be oper-
ated dry for a maximum of 60
seconds. Danger of overheating!
3.
Remove the sample return line from the pho-
tometers to the water tank.
4.
Remove the cover from the water tank.
Summary of Contents for AquaScat SG