Instruction Manual ScrubberGuard
Servicing
70
14830E/1
9.2
Servicing duties on the system
9.2.1
ScrubberGuard checklist
The following checklist is used for maintaining measuring operation. If something is found,
consult the corresponding section:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
General visual inspection of the pipes:
Are there indications of leaky screw con-
nections?
Are the pipes damaged?
2.
Visual inspection of the deaeration tubes:
Is there an excessive deposit build-up on
the glass of the venting tube?
Is there an absence of water/air flowing
through the respective venting tube?
Is the spout in the venting tube blocked
(circle)?
Are there leaky connections?
Are the venting tubes damaged?
If any of these questions are an-
swered in the affirmative, please
consult the Section 9.2.4.
3.
Visual inspection of the water tank:
Is the inside of the water tank dirty?
Are there indications of leaky connections?
Is the float switch inserted incorrectly (see
figure)?
Is there rust on the water tank?
If any of these questions are an-
swered in the affirmative, please
consult the Section 9.2.3.