Instruction Manual ScrubberGuard
Settings
62
14830E/1
8.3
Setting the limits
MANIPULATION
ADDITIONAL INFO / IMAGES
1.
Press the
Menu
button.
2.
Set the access code and confirm with
OK
.
Factory setting is
0
.
3.
Press the
Local ……
button.
If the desired menu does not
appear, press the arrow at the bot-
tom right.
4.
Press the
Limits
button.
5.
Select
Limit 1 .. 8
.
6.
Select the source of the measuring channel
from the
Source
menu item. This name is dis-
played to simplify identification of the measur-
ing channel.
The channels defined under
Meas.channels
, three math chan-
nels and two analog channels are
available for selection.
Reference
Manual
7.
Define the
Mode
.
The following selection is available:
Inactive
(limit monitoring of
this channel is deactivated)
Exceeded
(limit active when
the set threshold value is ex-
ceeded)
Undershot.
(limit active when
the set threshold value is under-
shot)
8.
Define the upper limit, lower limit, cut-in delay
and cut-out delay with the number pad.
Pressing the current number
value takes you to the entry mode.
9.
Press the
Meas
button.
The instrument is in measuring op-
eration again.
The limits have to be configured accordingly so that they are not only displayed, but that the
outputs are also switched.