General safety points
26
14830E/1
3.3.2
Calibration solutions
WARNING!
Incorrect measurement following the use of an incorrect or expired calibration solu-
tion for recalibration.
If an incorrect or expired calibration solution is used for recalibration, the measuring accura-
cy of the instrument may be affected. If recalibration cannot be made due to the loss of the
calibration solutions, the measuring accuracy of the instrument also cannot be guaranteed.
Always ensure that the correct calibration solution is used (e.g. pH 4).
Pay attention to the indicated expiry date and order a new calibration solution in good
time.
Always store the calibration solution according to the prescribed storage conditions (in a
dark place at room temperature).
Store the calibration solution in a defined location.
3.4
Residual risk
WARNING!
According to the risk assessment of the applied safety directive DIN EN 61010-1, the
following residual risks remain:
Overflow in the water circuit. This risk can be minimized with the following
measures:
Set up the system in locations where water drainage is guaranteed.
Use shut-off valves at the sampling point and sample return point.
Incorrect display of measuring values. This risk can be minimized with the following
measures:
Use an access code to prevent unauthorized persons from changing the parameters.
Perform the specified servicing duties.
In the event of faults, the surface of the pumps can become hot. The risk of injuries
can be minimized as follows:
Do not touch surfaces that are marked with the “Hot surface” warning symbol without
gloves.