Servicing
Instruction Manual ScrubberGuard
14830E/1
97
9.4.3
Manual adjustment
The following procedure describes how manual adjustment is made with an AquaScat SG /
OilGuard SG.
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Interrupt the sample feed to the photometer.
2.
Loosen the screw connections X and Y and
remove the inlet pipe.
3.
Remove the optics unit on the photometer
according to Section 9.4.1 and place in the
recess (Dockingstation).
Danger due to UV radia-
tion (on OilGuard SG only).
Use UV goggles and gloves.
4.
Check whether the appropriate checking unit
is used for the photometer and that the
checking unit is clean.
(OilGuard SG only)
The checking unit must have the
same temperature as the instru-
ment. If the checking unit is stored
in another room, let the unit accli-
matize to the conditions.
5.
Mount the checking unit and fasten with the
knurled screws.
Make sure that the pins enter the positioning
holes (arrows).
Example: OilGuard 2 SG