INSTRUCTION MANUAL AquaScat 2
Servicing
56
11778E/1
8.5.
Clean parts which come into contact with water
The following describes cleaning the parts which come into contact with water (Remember
however there are no windows to clean in the AquaScat WTM 2):
WORKSTEP
PROCEDURE
1.
Interrupt the sample sup-
ply to the photometer.
2.
Remove inlet hose and
drain hose from the pho-
tometer.
3.
Remove the optics unit of
the photometer and fas-
ten onto the docking sta-
tion.
Section 4.4
4.
Remove parts which come
into contact with water.
4.1: Loosen conduit gland (A) and remove the inlet
tube B).
4.2: Pull off the outlet cone (C) upward.