INSTRUCTION MANUAL AquaScat 2
Servicing
54
11778E/1
WORKSTEP
PROCEDURE
8.
Perform adjustment.
Press the
initiate
button and wait.
If the adjustment was successful, it is confirmed with
Adjustment OK
. This completes the adjustment.
If the adjustment was
not
successful, it is indicated with
Adjustment fault
. In this case, check the points in the
following list one after the other:
Cleanliness of the checking unit.
Correct checking unit used. (Compare serial No)
Nominal value does not correspond to the value
of the checking unit.
Soiled optics in the instrument.
In this case check the cleanliness of the optics as
described in Section 8.7 and then repeat the
procedure.
If the check could not be successfully completed,
contact your country representative.
Section 10
9.
Remove checking unit.
10. Mount the inlet tube as de-
scribed in Section 8.5, point
7, and assemble the instru-
ment in the reverse order.
11. The instrument can now be
operated again.
A new recalibration factor is determined during recalibration. The deviation from the original
state is displayed under
Curr. corr.
.