Servicing
INSTRUCTION MANUAL AquaScat 2
11778E/1
51
WORKSTEP
PROCEDURE
7.
Replace optics unit back onto
the photometer.
Replace the optics unit back onto the measuring
cell unit and lock with the four mounting clips. Pay
attention to the guide pins (see picture).
8.2.1.
Flow rate reduction
A reduction of the minimum flow rate from 2.5 l/min. to 1.3 l/min. is possible; however,
note the following points:
Reproducibility worsens (from ±1% at 2.5 l/min to ±3% at 1.3 l/min).
The instrument must be very precisely leveled. If there is a 1° angle, the measuring val-
ue can change by up to ±3%.
If the flow rate is low, it is even more important to maintain a constant supply flow be-
cause the fluctuations have a greater impact on the measuring value.