INSTRUCTION MANUAL AquaScat 2
Initial start-up
28
11778E/1
6.
Initial start-up
The initial start-up with the web user interface via the Ethernet interface is described in the
Reference Manual.
Proceed with the initial start-up in accordance with the following table. If malfunctions oc-
cur, please refer to Section 9.
WORKSTEP
PROCEDURE
1.
Remove the optics unit of the
photometer and fasten onto the
docking station.
Section 4.4
2.
Check the sampling system for
the correct sample feed and
then open the sample feeding.
Section 4.5
Flooding of the instrument due to im-
proper mounting of the drain hose:
The drain hose must be fastened so that the flood-
ing protection holes on the side are not covered
(flood protection).
3.
Check the flow rate and ensure
that the light trap and aperture
are clean.
There must be no pres-
sure surge when turning the
water off and on!
Section 8.2
4.
Make sure that the photometer
is correctly mounted and that
the electrical connections are
correctly connected.
Section 4.2