Mounting the measuring device
4.1.
Location evaluation
A note about selecting the right installation location:
The sample feed should be as short as possible so that changes in the reading are dis-
played without delay.
There should be no vibrations on the mounting surface. Vibrations cause measuring er-
rors and can permanently damage the unit.
The photometer and its peripherals should be easily accessible to facilitate inspection
and servicing duty.
Make sure that the photometer is not exposed to direct sunlight. This may cause signifi-
cant deviations in the reading.
4.2.
Mechanical mounting of the photometer
The dimensional specifications according to the mounting drawings in the Appendix must be
observed when mounting the photometer.
Section 15
WORKSTEP
PROCEDURE
1. Fit the mounting bracket.
Screw on the mounting bracket at the intended position.
The mounting bracket must be horizontally aligned with
a spirit level.