INSTRUCTION MANUAL AquaScat 2
Operation
46
11778E/1
MANIPULATION
SUPPLEMENTARY INFORMATION
6.
Activate output functions (multiple selection
possible)
Activated output functions are high-
lighted green.
Invert: inverts the outputs
Prio fault
Fault
Warning
Service
Adjustment
Limit 1.2
The functions marked with MR refer to
automatic measuring range switching
and are described in the Reference
Manual.
7.
Press the
Meas
button.
Instrument again in normal operation.
7.12.
Setting the date and time
MANIPULATION
SUPPLEMENTARY
INFORMATION
1.
Press the
Menu
button.
2.
Set the access code and confirm with
OK
.
Factory setting is
0
.
3.
Use the
down
arrow key to select the second page.
4.
Press the
Configuration
button.
5.
Press the
down
arrow key.
6.
To enter the time, press the currently displayed time at
the
Time
menu item and enter the new time with the
number pad.
Confirm entry with
OK
.
The time must be entered
in the format
hh:mm
.
7.
To enter the date, press the currently displayed date at
the
Date
menu item and enter the new date with the
number pad.
Confirm entry with
OK
.
The date must be entered
in the format
DD.MM.YYYY.
8.
Press the
Meas
button.
Instrument again in normal
operation.