INSTRUCTION MANUAL AquaScat 2
Mounting the measuring device
20
11778E/1
WORKSTEP
PROCEDURE
3.
Fasten drain hose onto the photom-
eter.
Flooding of the instrument
due to improper mounting of the
drain hose:
The drain hose must be fastened so
that the flooding protection holes
on the side are not covered (flood
protection).
Section 4.5.1
Fasten the drain hose on the drain tube (Figure
4, 7) of the instrument.
When doing so, press the outlet tundish (Figure
4, 5) downward with your hand so that the
drain tube (Figure 4, 7) is not pressed into the
measuring cell space. This can interfere with
the operation of the automatic calibration sys-
tem so care is required.
4.
Check the position of the inlet tube
and outlet tundish.
4.1: Check to ensure the inlet tube (Figure 4, 1)
projects 5 mm out of the holder.
4.2: Check whether the outlet tundish (Figure
4, 5) contacts the conduit gland (Figure 4, 6).
5.
Place the optics unit again on the
measuring cell unit and lock with
the four mounting clips.