Operation
INSTRUCTION MANUAL AquaScat 2
11778E/1
37
7.5.
Switching to service operation
The photometer is configured in the service mode. The measuring procedure is suspended
and the main menus appear on the display.
Proceed as follows for service operation:
MANIPULATION
SUPPLEMENTARY
INFORMATION
1.
Press the
Menu
button.
2.
Set the access code and confirm with
OK
.
Factory setting is
0
.
3.
The main menus appear.
Now you are in service operation.
The following applies in service operation:
* The measuring values remain on the last values on the digital interfaces.
* Depending on the configuration, the current outputs go to 0/4 mA or remain on the
last measuring value.
The limits will be deactivated.
If an output for service is programmed, it is activated.
Error messages are suppressed.
* This doesn't apply when the
Current outputs\General\For service
parameter is set to
Measure
.
For normal operation press the
Meas
button. When switching from service operation to
normal operation, an hourglass appears in the information bar for about 20 seconds. The
measuring values are frozen during this time.
Definition
Service mode