Operation
INSTRUCTION MANUAL AquaScat 2
11778E/1
45
7.10.
Configuring the flow meter with limit switch
The optional flow meter with limit switch is connected to terminals 22, 23 (In1). Proceed as
follows to activate monitoring:
MANIPULATION
SUPPLEMENTARY INFORMATION
1.
Press the
Menu
button.
2.
Set the access code and confirm with
OK
.
Factory setting is
0
.
3.
Press the
Inp./outputs
button.
4.
Press the
Inputs
button.
5.
Select
Input 1
.
6.
Activate input functions
External
and
In-
vert
.
7.
Press the
OK
button.
8.
Press the
General
button.
9.
Under
Name ext.in
you can enter the text
of the error message with a maximum of
seven characters (e.g. FLOWRAT).
Prio.ext.in
should be on
Fault
.
10.
Press the
Meas
button.
Instrument again in normal operation.
7.11.
Setting the outputs (relay outputs 1 or 2)
Outputs 1/2 are located are on the base print of the customer connection terminals and are
designated as relays 1/2.
Section 5.3
MANIPULATION
SUPPLEMENTARY INFORMATION
1.
Press the
Menu
button.
2.
Set the access code and confirm with
OK
.
Factory setting is
0
.
3.
Press the
Inp./outputs
button.
4.
Press the
Outputs
button.
5.
Select output 1 or 2 button.