INSTRUCTION MANUAL AquaScat 2
Troubleshooting
68
11778E/1
9.
Troubleshooting
9.1.
Pinpointing the cause of a malfunction
DETECTABLE MALFUNCTION MEASURES
No reading.
Check whether the service voltage is connected.
Section 5
Check whether the fine-wire fuse in the instrument
is OK.
Section 9.2
Fault report in the display.
Analyze the error message.
Section 9.3
The measuring value appears
wrong.
Make sure there are no air bubbles in the sample.
Section 2.1 (deaeration tube)
Ensure that the sample to be measured corresponds
to the operating conditions.
Section 2.4
Perform adjustment.
Check whether the photometer is correctly mount-
ed.
Section 4
Ensure that the servicing duties have been per-
formed according to the servicing schedule.
Section 8.1
Clean the optics.
Section 8.7
If the listed measures do not result in the desired results, please consult with customer ser-
vice.
Section 10