Servicing
INSTRUCTION MANUAL AquaScat 2
11778E/1
55
8.4.
Automatic adjustment for AquaScat 2 WTM A
The automatic adjustment is possible only with instrument type AquaScat 2 WTM A.
The automatic adjustment can also be set for timed actuation in the
Adjust interval
menu. This is described in the Reference Manual.
The following describes the manual initiation of the automatic adjustment:
MANIPULATION
SUPPLEMENTARY
INFORMATION
1.
Switch photometer to service operation.
Section 7.5
2.
Press the
Recalibration
and then the
C1 Turb
but-
ton.
3.
Press the
initiate
button. This starts the adjustment.
If the adjustment was successful, it is confirmed with
Adjustment OK
. This completes the adjustment.
If the calibration check was
not
successful, it is indi-
cated with
Adjustment fault
. The following points
can be checked:
Check the cleanliness of the optics as described
in Section 8.7.
Check whether the outlet cone is correctly
mounted according to Section 8.5, point 6.
After performing the checks above, repeat the ad-
justment.
If the adjustment could not be successfully
completed, contact your country representative.
Section 10
A new recalibration factor is determined during the adjustment. The deviation from the orig-
inal state is displayed under
Curr. corr.
.