3 Safety
Version 05/2010, Rev. 2.12 of 15/08/2016 • sb
21 / 70
Translation of the original operating manual, part no. 0700102
Pos: 57 / 100 Sigma /100 BA Ze ntrifu gen Sigma (Stan da rdm odule )/0 30 Sich erh eit/03 0-0 060 -00 70 Sicher heit v on Ro tor en u nd Zub ehö r @ 44\ mod_ 147 011 920 056 8_6 8.doc x @ 4 370 08 @ 3 @ 1
3.6.7 Safety of rotors and accessories
Pos: 58 / 100 Sigma /100 BA Ze ntrifu gen Sigma (Stan da rdm odule )/0 30 Sich erh eit/03 0-0 060 -00 70- 001 0 Lebe nsd aue r @ 2 6\m od_ 140 531 9032 612 _68 .docx @ 1 9214 3 @ 3 @ 1
3.6.7.1 Service life
The rotors and accessories have a limited service life.
WARNING
• Perform regular checks (at least once per month) for safety reasons!
• Pay special attention to changes, such as corrosion, cracks, material
abrasion, etc.
• After 10 years, they must be inspected by the manufacturer.
• After 50,000 cycles, the rotor must be scrapped for reasons of safety.
• If other data concerning the service life are engraved on the rotor or
bucket, these data shall apply accordingly. For example, a bucket with
the engraving “max. cycles = 10,000” has a service life of 10,000
cycles, and a rotor with the engraving “Exp. date 02/20” must be
scrapped in February 2020 at the latest (see figure).
Fig. 4: Different service life – engraving on the bucket/rotor
NOTE
• Refer to the table of rotors and accessories with a different service
life (see chapter 11.3 - "Table of the service life of rotors and
accessories")!