6.3.7 Runtime
The preselected total runtime is displayed in the lower line of the display.
During centrifugation, the remaining runtime is displayed. The runtime of
the centrifuge can be set at one-second-intervals up to 99 minutes and 59
seconds.
• To select a runtime, press the set key repeatedly until the time unit
flashes on the display.
• Select the desired runtime by pressing the arrow keys.
• Confirm the selection by pressing the start/stop key.
The data will also be accepted and saved after approx. 20 seconds if no
other key is pressed within this time period.
Pos: 127 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1
Pos: 128 /10 0 Sigm a/10 0 BA Z entrif uge n Sigm a (Sta nda rd mod ule)/ 061 B etrie b_Spinc ont rol Basic/Spi ncon trol Ba sic /06 1-0 030 -00 70- 001 0 Änd ern de r La ufzeit w ähr end des Lau fs @ 2 7\m od_ 140 532 035 828 2_68 .docx @ 1 938 79 @ 4 @ 1
6.3.7.1 Changing the runtime during centrifugation
The preselected runtime can be changed during centrifugation.
• Press the "set" key repeatedly until the time unit flashes on the display.
• Change the desired runtime by pressing the arrow keys. The
parameters will be accepted immediately.
NOTE
If the centrifugation time is changed during the run, the centrifuge will run
for the entire new time and will disregard the previous runtime that has
already elapsed.
Pos: 129 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1
Pos: 130 /10 0 Sigm a/10 0 BA Z entrif uge n Sigm a (Sta nda rd mod ule)/ 061 B etrie b_Spinc ont rol Basic/Spi ncon trol Ba sic /06 1-0 030 -00 70- 002 0 Kurz zeitb etrie b ( nur 1-1 4, 1 -14K ) @ 2 8\m od_ 140 532 1080 308 _68 .docx @ 1 9609 1 @ 4 @ 1
6.3.7.2 Short run (quick run)
During the short run, the centrifuge accelerates at maximum power until the
maximum speed is reached.
• Keep the start/stop key pressed during the short run.
The message "short run" and the duration of the short run is displayed.
When the start/stop key is released, the centrifuge decelerates at maximum
power to a standstill.
After the short run, the lid unlocks automatically and the program that was
set beforehand is displayed again.
Pos: 131 /01 0 Univ ersal mod ule/Seit enwec hsel @ 0\m od_ 120 211 624 4312 _0. docx @ 10 5 @ @ 1