6 Using the centrifuge
Version 05/2010, Rev. 2.12 of 15/08/2016 • sb
31 / 70
Translation of the original operating manual, part no. 0700102
Pos: 107 /10 0 Sigm a/10 0 BA Z entrif uge n Sigm a (Sta nda rd mod ule)/ 061 B etrie b_Spinc ont rol Basic/ 061 -00 20- 002 0-0 030 Ein setze n vo n Z ube hör @ 27 \mo d_1 405 320 3186 88_ 68. docx @ 19 3739 @ 4 @ 1
6.2.2.4 Installation of accessories
• Only use tubes that are suitable for the rotor (see chapter 11.1 -
"Range of accessories").
• Always load the axial symmetrical inserts of the rotors with the same
accessories and fill to avoid imbalance.
Centrifugation with low capacity
• Install the tubes axial symmetrically so that the rotor is loaded evenly.
• It is not permissible to load angle rotors on only one axis.
correct
incorrect
Fig. 11: Permissible and impermissible loading of an angle rotor (example illustration)
Pos: 108 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1
Pos: 109 /10 0 Sigm a/10 0 BA Z entrif uge n Sigm a (Sta nda rd mod ule)/ 061 B etrie b_Spinc ont rol Basic/ 061 -00 20- 002 0-0 040 Gefä ße @ 27 \mo d_1 405 320 320 154_ 68. docx @ 19 375 3 @ 4 @ 1
6.2.2.5 Tubes
• Load the tubes outside of the centrifuge. Liquids in the bores of the rotor
cause corrosion.
• Fill the tubes carefully and arrange them according to their weight.
Imbalances result in the excessive wear of the bearings.
• Follow the safety instructions and hazard warnings (see chapter 3 -
"Safety")!