Refrigerated Centrifuge Sigma 1-14K
8 Maintenance and service
Version 05/2010, Rev. 2.12 of 15/08/2016 • sb
45 / 70
Translation of the original operating manual, part no. 0700102
Pos: 172 /10 0 Sigm a/10 0 BA Z entrif uge n Sigm a (Sta nda rd mod ule)/ 080 W art ung und Inst andh altun g/0 80- 001 0-0 020 -00 10 Ku nstst offzu beh ör @ 28\ mod _14 053 206 1960 2_6 8.d ocx @ 194 859 @ 4 @ 1
8.1.3.1 Plastic accessories
The chemical resistance of plastic decreases with rising temperatures (see
chapter 11.4 - "Resistance data").
• If solvents, acids, or alkaline solutions have been used, clean the plastic
accessories thoroughly.
WARNING
Plastic accessories must not be greased!
Pos: 173 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1
Pos: 174 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1
Pos: 175 /10 0 Sigm a/10 0 BA Z entrif uge n Sigm a (Sta nda rd mod ule)/ 080 W art ung und Inst andh altun g/0 80- 001 0-0 031 Rot ore n (n ur Basic ) @ 28\ mod _14 0532 062 233 6_6 8.do cx @ 1 948 87 @ 3 @ 1
8.1.4 Rotors
The rotors are produced with the highest precision, in order to withstand the
permanent high stress from high gravitational fields.
Chemical reactions as well as stress-corrosion (combination of oscillating
pressure and chemical reaction) can affect or destroy the metals. Barely
detectable cracks on the surface can expand and weaken the material
without any visible signs.
• Check the material regularly (at least once a month) for
– cracks
– visible damage of the surface
– pressure marks
– signs of corrosion
– other changes.
• Check the bores of the rotors and multiple carriers.
• Replace any damaged components immediately for your own safety.
Pos: 176 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1
Pos: 177 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1
Pos: 178 /10 0 Sigm a/11 0 BA Z entrif uge n Sigm a (Pr ojekt e)/1 -14 _1- 14K/0 80 Wa rtu ng u nd I nstan dhalt ung /08 0-0 010 -00 40 Mi kro häm atok rittelle r 1 -14 _1- 14K @ 2 8\m od_ 140 532 086 2886 _68 .docx @ 1 956 85 @ 3 @ 1
8.1.5 Microhaematocrit rotor
• Remove the microhaematrocrit rotor for cleaning.
• Wipe the centrifuge chamber clean.
• Replace the rubber ring in the event of wear or glass breakage (part no.
16002 for rotor 11026).