8 Maintenance and service
The centrifuge, rotor, and accessories are subject to high mechanical
stress. Thorough maintenance performed by the user extends the service
life and prevents premature failure.
CAUTION
If corrosion or other damage occurs due to improper care, the
manufacturer cannot be held liable or subject to any warranty claims.
• Use soap water or other water-soluble, mild cleaning agents with a pH
value between 6 and 8 for cleaning the centrifuge and accessories (see
also chapter 8.2 - "Sterilisation and disinfection of the rotor chamber
and accessories").
• Avoid corrosive and aggressive substances.
• Do not use solvents.
• Do not use agents with abrasive particles.
• Do not expose the centrifuge and rotors to intensive UV radiation or
thermal stress (e.g. by heat generators).
Pos: 163 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1
Pos: 164 /10 0 Sigm a/10 0 BA Z entrif uge n Sigm a (Sta nda rd mod ule)/ 080 W art ung und Inst andh altun g/0 80- 001 0 Wa rtun gsa rbeit en- --- -- --- -- --- -- --- --- -- --- -- --- --- -- -- @ 28\ mod _14 053 206 139 78_ 68.d ocx @ 194 803 @ 2 @ 1
8.1 Maintenance
Pos: 165 /10 0 Sigm a/10 0 BA Z entrif uge n Sigm a (Sta nda rd mod ule)/ 080 W art ung und Inst andh altun g/0 80- 001 0-0 010 Z entri fuge @ 2 8\m od_ 1405 320 615 295 _68. docx @ 19 481 7 @ 3 @ 1
8.1.1 Centrifuge
• Unplug the mains power plug before cleaning.
• Carefully remove all liquids, including water and particularly all the
solvents, acids, and alkaline solutions from the rotor chamber using a
cloth in order to avoid damage to the motor bearings.
• If the centrifuge has been contaminated with toxic, radioactive, or
pathogenic substances, clean the rotor chamber immediately with a
suitable decontamination agent (depending on the type of
contamination).
WARNING
Take suitable precautions for your own safety if there is a risk of toxic,
radioactive, or pathogenic contamination.
• Grease the motor shaft slightly after cleaning (grease for load-bearing
bolts part no. 70284).