Refrigerated Centrifuge Sigma 1-14K
5 Set-up and connection
Version 05/2010, Rev. 2.12 of 15/08/2016 • sb
25 / 70
Translation of the original operating manual, part no. 0700102
Pos: 84 / 100 Sigma /100 BA Ze ntrifu gen Sigma (Stan da rdm odule )/0 50 Aufs tellun g u nd Ansc hluss/0 50 A ufstellu ng und Ans chluss == === == == === == == == === == == === == == == @ 27\ mod _14 053 191 730 71_ 68.d ocx @ 192 381 @ 1 @ 1
5 Set-up and connection
Pos: 85 / 100 Sigma /100 BA Ze ntrifu gen Sigma (Stan da rdm odule )/0 50 Aufs tellun g u nd Ansc hluss/0 50 -001 0 Au fstello rt ( auß er 8KS) @ 2 7\m od_ 140 531 917 4276 _68 .docx @ 1 923 95 @ 2 @ 1
5.1 Installation site
Operate the centrifuge only in closed and dry rooms.
All the energy supplied to the centrifuge is converted into heat and emitted
to the ambient air.
• Ensure sufficient ventilation.
• Keep a safety distance of at least 30 cm around the centrifuge so that
the vents in the centrifuge remain fully effective.
• Do not subject the centrifuge to thermal stress, e.g. by positioning it
near heat generators.
• Avoid direct sunlight (UV radiation).
• The table must be stable and have a solid, even surface.
• Attention: During transport from cold to warmer places, condensational
water will collect inside the centrifuge. It is important to allow sufficient
time for drying (min. 24 h) before the centrifuge can be used again.
Pos: 86 / 010 Unive rsalm odul e/ L ee rzeile @ 0\ mod _120 211 624 450 0_0. docx @ 11 4 @ @ 1
Pos: 87 / 100 Sigma /100 BA Ze ntrifu gen Sigma (Stan da rdm odule )/0 50 Aufs tellun g u nd Ansc hluss/0 50 -002 0 En ergi evers or gung -- --- -- --- --- -- --- -- --- -- --- --- -- --- -- --- --- -- --- -- --- @ 2 7\mo d_1 405 319 175 630 _68. docx @ 19 240 9 @ 2 @ 1
5.2 Power supply
Pos: 88 / 100 Sigma /100 BA Ze ntrifu gen Sigma (Stan da rdm odule )/0 50 Aufs tellun g u nd Ansc hluss/0 50 -002 0-0 010 Anschl ussa rt (Kalt ger ätes tecke r) @ 32\ mo d_14 175 996 706 21_ 68.d ocx @ 222 346 @ 3 @ 1
5.2.1 Type of connection
DANGER
The operating voltage on the name plate must correspond to the local
supply voltage!
Sigma centrifuges are units of protection class I. The centrifuges of this
model series have a three-wire power cord with an IEC C13 connector.
Pos: 89 / 010 Unive rsalm odul e/ L ee rzeile @ 0\ mod _120 211 624 450 0_0. docx @ 11 4 @ @ 1
Pos: 90 / 100 Sigma /100 BA Ze ntrifu gen Sigma (Stan da rdm odule )/0 50 Aufs tellun g u nd Ansc hluss/0 50 -002 0-0 020 Siche run gen baus eits (au ßer 3- 30KS) @ 7\m od_ 130 6312 302 397 _68 .docx @ 43 513 @ 3 @ 1
5.2.2 Customer-provided fuses
Typically, the centrifuge must be protected with 16 Amp L or B fuses that
are to be provided by the customer.
Pos: 91 / 010 Unive rsalm odul e/Absch nittswec hsel @ 0\m od_ 120 212 451 406 2_0. docx @ 41 8 @ @ 1
Pos: 92 / 010 Unive rsalm odul e/Seite nwechs el @ 0 \mo d_1 202 116 244 312_ 0.d ocx @ 105 @ @ 1