6. Installation
The Transducer should be mounted in a reasonably stable ambient
temperature and where the operating temperature is within the range
0
0 to 45 C. Vibration should be kept to a minimum and the product should
not be mounted where it will be subjected to excessive direct sunlight.
As the front of the enclosure conforms to IP 40. The terminals of the
product should be protected from liquids.
Transducer
by
can be mounted either on a top-hat rail or directly
on to a wall a mounting plate.
Caution
1.
In the interest of safety and functionality this product must be installed by
a qualified engineer, abiding by any local regulations.
2.
Voltages dangerous to human life are present at some of the terminal
connections of this unit. Ensure that all supplies are de-energised before
attempting any connection or disconnection.
3.
These products do not have internal fuses therefore external fuses must
be used to ensure safety under fault conditions.